What actually needs to be calibrated
Salt spray chamber calibration covers a handful of specific parameters, not the chamber as a single unit:
- Chamber temperature — verified against an independent calibrated thermometer or data logger, not just the chamber's built-in display, since controller drift is common after months of continuous salt exposure.
- Fog collection rate — checked using graduated collection jars placed inside the chamber over a fixed period; ASTM B117 and ISO 9227 specify an acceptable collection rate range.
- Salt solution concentration and pH — verified against the standard's specified range (typically 5% NaCl, pH 6.5–7.2 for neutral salt spray) using a hydrometer or refractometer and a pH meter.
- Compressed air pressure to the atomizer, where applicable — affects droplet size and fog consistency.